Automatic Flammability Test Chamber (50W/500W) – QualiFlame™ UL94 A
Automatic Horizontal and Vertical Burning Test Chamber for UL 94 Classification
The QualiFlame™ UL94 A is an automated flammability testing system designed to evaluate the burning characteristics of plastics, printed circuit boards, and completed electrical components. It accommodates both horizontal and vertical test procedures, supporting HB, V-0, V-1, V-2, 5VA, and 5VB classifications. Qualitest offers two burner options, rated at 50 W and 500 W, to cover the required test methods. Its stainless steel enclosure and internal construction provide resistance to smoke deposits and corrosive gases produced during combustion. A PLC-based control system with a touchscreen coordinates ignition, flame exposure, and test operations.
The testing chamber and control section are housed within a single cabinet, reducing the complexity of equipment installation and commissioning. With an internal volume exceeding 0.5 m³, the chamber provides sufficient observation space for testing, while its black interior improves the visibility of small flames and residual glowing. Operators can configure flame exposure duration, afterflame timing, and burner positioning directly through the control panel. The burner assembly offers three angular settings (0°, 20°, and 45°), allowing the required orientations for horizontal, vertical, and inclined specimen testing.
Accurate flame verification is performed using a drilled copper calibration block fitted with a 0.5 mm Type K thermocouple. During this procedure, the block must reach a specified temperature increase within a predetermined period, confirming that the burner produces the required thermal output before a test proceeds. The burner operates on methane with a minimum purity of 98%. Its gas flow is regulated to 965 ml/min for the 500 W configuration and 105 ml/min for the 50 W configuration. Both burner versions feature a tube with a 9.5 mm internal bore.
Automatic Flammability Test Chamber (50W/500W) – QualiFlame™ UL94 A Applications
The QualiFlame™ UL94 A is suitable for production facilities, material development departments, and testing laboratories that need to establish the fire performance of plastic materials and electrical components. By reproducing controlled horizontal and vertical flame exposure conditions, the system helps determine whether materials meet the required UL 94 classifications at different stages of manufacturing, product development, and quality assurance.
-
Polymer Formulation and Material Development: Supports the evaluation of newly formulated resins and flame-retardant compounds by determining their burning characteristics and assigning the appropriate UL 94 classification. Test results help assess polymer grades and flame-retardant packages during formulation development.
-
Printed Circuit Board and Electronic Substrate Manufacturing: Provides burning performance assessments for PCB substrates and laminate materials, helping manufacturers verify whether their materials achieve the V-0 classification commonly specified for electronic assemblies.
-
Electrical Hardware and Component Production: Helps manufacturers establish the flammability performance of plastic parts used in connectors, terminal blocks, switches, and relay housings. These evaluations provide classification data before components are submitted to certification agencies.
-
Household Appliance Manufacturing: Evaluates the fire behavior of plastic appliance enclosures and internal structural supports. Testing helps determine whether these materials satisfy the flammability requirements referenced by applicable appliance safety standards.
-
Cable Management and Electrical Installation Products: Examines how plastic accessories respond to flame exposure, including whether they continue burning or produce flaming droplets. Typical specimens include cable ties, cable glands, conduits, and junction boxes.
-
Automotive and Electric Vehicle Manufacturing: Assesses plastic materials used in battery pack enclosures, busbar supports, and charging connectors where higher flammability classifications, particularly 5VA and 5VB, are required.
-
Lighting Equipment and Luminaire Production: Determines the burning characteristics of plastic diffusers, driver housings, and lamp holders that operate in environments where components may experience sustained heat exposure.
-
Raw Material and Component Quality Assurance: Enables quality control teams to compare the flammability performance of delivered polymer compounds and finished molded parts against the classifications declared in their respective datasheets.
-
Product Certification and Independent Testing Services: Enables third-party laboratories and certification facilities to conduct standardized horizontal and vertical burning evaluations. The resulting test data supports material approval assessments and the preparation of product listing documentation.
Standards
The QualiFlame™ UL94 A accommodates the prescribed test flame configurations and specimen evaluation procedures used to determine the fire performance classifications of plastic materials. Its testing capabilities cover the following international and product safety standards:
-
IEC 60695-11-3 – Fire Hazard Testing, Part 11-3: Test Flames, 500 W Flames, Apparatus and Confirmational Test Methods.
-
IEC 60695-11-4 – Fire Hazard Testing, Part 11-4: Test Flames, 50 W Flame, Apparatus and Confirmational Test Method.
-
IEC 60695-11-10 – Fire Hazard Testing, Part 11-10: Test Flames, 50 W Horizontal and Vertical Flame Test Methods.
-
IEC 60695-11-20 – Fire Hazard Testing, Part 11-20: Test Flames, 500 W Flame Test Methods.
-
UL 94 – Standard for Tests for Flammability of Plastic Materials for Parts in Devices and Appliances.
Automatic Flammability Test Chamber (50W/500W) – QualiFlame™ UL94 A Key Features
Designed for repeatable horizontal and vertical burning evaluations, the QualiFlame™ UL94 A integrates automated test control, adjustable burner positioning, and dedicated flame calibration equipment. These features support consistent testing conditions and provide the verification needed for reliable material classification.
-
Two Burner Options for 50 W and 500 W Testing: Provides the flame configurations required for HB, V-series, and 5V classifications within one testing system, eliminating the need to transfer specimens between separate instruments.
-
PLC-Based Automated Operation: Coordinates flame application, exposure periods, and burner withdrawal through programmed sequences, minimizing variations caused by manual timing and operator response.
-
Three-Position Burner Adjustment: Allows selection of 0°, 20°, or 45° burner angles to accommodate the specimen orientations required for horizontal, vertical, and inclined burning procedures.
-
Corrosion-Resistant Stainless Steel Construction: Features a stainless steel chamber enclosure and internal components designed to withstand accumulated smoke residue and corrosive combustion byproducts during repeated testing.
-
Combined Testing and Control Cabinet: Integrates the operating controls and test chamber into one assembly, simplifying installation and reducing commissioning requirements at the testing location.
-
Large Chamber with Black Interior: Provides an internal volume greater than 0.5 m³ and a dark observation background, making low-intensity flames and afterglow easier to identify during testing.
-
Copper Block Calibration System: Uses a drilled copper block to check burner performance against the required temperature increase and heating interval before specimens are evaluated.
-
Fine-Gauge Type K Thermocouple: Incorporates a 0.5 mm thermocouple for responsive temperature measurement during the flame calibration procedure, supporting accurate monitoring of the timed temperature rise.
-
Controlled Methane Gas Delivery: Maintains the specified flow rates of 965 ml/min for the 500 W burner and 105 ml/min for the 50 W burner to provide the corresponding flame energy.
-
Flame Height Reference Indicator: Helps operators verify the appropriate 40 mm or 20 mm flame height at the burner before beginning each series of tests.
-
Standardized 9.5 mm Burner Bore: Uses a burner tube with a 9.5 mm internal diameter to maintain the geometry specified for the test flame and support consistent flame formation.
-
English-Language Display with Customization Options: Includes an English operating interface suitable for North American laboratories, with alternative language configurations available upon request.
-
Adaptable Electrical Supply: Offers power supply customization to accommodate regional mains requirements beyond the standard 220 V single-phase configuration.
Theory and Method
UL 94 flammability ratings are established by observing how a plastic specimen responds to controlled flame exposure rather than measuring an isolated material characteristic. The classification process considers specimen thickness, flame intensity, exposure duration, burning rate, afterflame, afterglow, and flaming droplets. Each test method applies specific conditions to determine whether the material meets the corresponding performance requirements.
Horizontal Burning Test (HB)
For horizontal evaluation, the specimen is positioned flat and exposed to the 50 W test flame for 30 seconds. Once the flame has been applied, the operator measures how long the burning front takes to travel between two marked reference points. This recorded travel time is used to calculate the material's linear burning rate. The specimen receives an HB classification when it extinguishes before reaching the first reference mark or when its burning rate remains below the applicable maximum limit.
Vertical Burning Test (V-0, V-1, and V-2)
Vertical testing introduces more demanding conditions because the rising flame can preheat the unburned portion of the specimen and influence subsequent combustion. The test specimen is suspended vertically above a cotton indicator layer and exposed to the 50 W flame twice, with each application lasting 10 seconds. Timing measurements capture the afterflame period following each flame application, along with the afterglow duration after the second exposure.
The V-0, V-1, and V-2 classifications are determined by the permitted afterflame and afterglow durations, together with the behavior of flaming droplets. A critical distinction between these ratings is whether burning material falling from the specimen ignites the cotton indicator positioned underneath. Consequently, two materials with otherwise comparable burning characteristics may receive different classifications because of their dripping behavior.
500 W Flame Testing for 5VA and 5VB
The 500 W test introduces substantially greater thermal energy, approximately an order of magnitude higher than the 50 W flame, and is used for 5VA and 5VB classification. During this procedure, specimens in bar and plaque forms undergo five flame applications, each lasting five seconds, with five-second intervals between applications.
In addition to observing the burning response, plaque specimens are examined for burn-through. Whether the flame penetrates the plaque is the determining factor that distinguishes the 5VA classification from 5VB.
Copper Block Flame Calibration
Reliable classification depends on maintaining the specified burner output before specimen testing begins. The drilled copper block provides a controlled reference for checking the thermal performance of the flame. During calibration, the block must achieve the prescribed temperature increase within the defined time limit. If the flame has not been properly calibrated, the resulting test comparison is invalid. This verification step therefore establishes the required flame conditions before horizontal, vertical, or 500 W burning evaluations are carried out.
Automatic Flammability Test Chamber (50W/500W) - QualiFlame™ UL94 A Technical Specifications
| Parameter |
500 W Configuration (5V Class) |
50 W Configuration (HB/V Class) |
| Test Flame |
500 W nominal |
50 W nominal |
| Classification Coverage |
5VA, 5VB |
HB, V-0, V-1, V-2 |
| Burner Tube Internal Diameter |
Ø9.5 mm ±0.3 mm |
Ø9.5 mm ±0.3 mm |
| Burner Angle |
0°, 20°, 45° |
0°, 20°, 45° |
| Thermocouple |
Type K (Ni/Cr-Ni/Al) |
Type K (Ni/Cr-Ni/Al) |
| Thermocouple Size |
0.5 mm |
0.5 mm |
| Calibration Copper Block |
Ø9 mm ±0.01 mm, 10 g ±0.05 g before drilling |
Ø5.5 mm ±0.01 mm, 1.76 g ±0.01 g before drilling |
| Temperature Rise Range |
100°C ±5°C to 700°C ±3°C |
100°C ±5°C to 700°C ±3°C |
| Temperature Rise Time |
54 s ±2 s (preset) |
44 s ±2 s (preset) |
| Gas Flow |
965 ±30 ml/min |
105 ±5 ml/min |
| Gas Supply |
Methane, 98% minimum purity |
Methane, 98% minimum purity |
| Water Column |
125 mm ±5 mm |
<10 mm |
| Test Time |
54 s ±2 s |
44 s ±2 s |
| Flame Height Indicator |
40 mm ±1 mm |
20 mm ±1 mm |
| Chamber Volume |
>0.5 m³, black interior |
>0.5 m³, black interior |
| Controller |
PLC with integrated touch interface |
PLC with integrated touch interface |
| Voltage |
AC 220 V, 50 Hz to 60 Hz, 800 VA (110V is also available) |
AC 220 V, 50 Hz to 60 Hz, 800 VA (110V is also available) |
| Display Panel Language |
English (other languages available on request) |
English (other languages available on request) |