Needle Flame Tester – QualiFlame™ NFT
Needle Flame Test Apparatus for Fire Hazard Testing
The QualiFlame™ NFT is designed to determine how electrical components and insulating materials behave when exposed
to a localized flame caused by an electrical fault. It generates a controlled 12 mm flame using a stainless steel
burner with a 0.5 mm bore, supplied with propane or butane. The tester is suitable for manufacturers of lighting
equipment, low-voltage switchgear, household appliances, electric motors, power tools, and electronic instruments.
It also supports testing activities in research facilities, manufacturing environments, and quality control
laboratories. An integrated PLC system and touchscreen interface manage the testing process.
Testing is conducted in accordance with the needle-flame procedure specified in IEC 60695-11-5. The burner offers
three positioning angles: 0°, 20°, and 45°, allowing the flame to contact the specimen in the orientation required
by the applicable product standard. Flame exposure can be programmed from 0 s to 999.9 s, with increments of 0.1 s.
Although the equipment supports this full timing range, a 30-second exposure is commonly specified by product
standards. Afterflame duration is captured automatically, while the operator can also stop the measurement manually
when necessary.
Before specimen testing, burner performance is checked using a Cu-ETP copper block and a 0.5 mm Type K thermocouple.
The calibration procedure requires the copper block to reach the specified temperature rise within 23.5 seconds,
providing confirmation that the flame meets the required thermal output. A test chamber with an internal volume
greater than 0.5 m³ supplies the draught-free environment required by the method. The system operates with
commercially available bottled lighter gas containing at least 95% propane or butane, eliminating the need for a
dedicated specialty gas supply.
Needle Flame Tester – QualiFlame™ NFT Applications
The QualiFlame™ NFT supports fire hazard assessments across electrical product manufacturing, material development,
and laboratory inspection. It helps determine whether components and insulating materials can withstand localized
flame exposure under the conditions specified by relevant safety requirements.
-
Lighting Products and Luminaires: Examines lamp holders, driver enclosures, and insulating
supports to determine their response when an internal electrical fault produces a flame.
-
Low-Voltage Switchgear and Controlgear: Assesses protective barriers, arc chutes, and molded
enclosures used in circuit breakers, contactors, and electrical distribution assemblies.
-
Household Appliance Manufacturing: Evaluates internal plastic parts and wiring supports
against needle-flame requirements referenced in appliance safety standards.
-
Electric Motors and Rotating Machines: Investigates the flame resistance of slot liners,
terminal boxes, and winding insulation systems under potential electrical fault conditions.
-
Power Tools and Portable Equipment: Tests handle materials, switch housings, and internal
supporting components that may encounter flames resulting from motor or contact faults.
-
Electrical and Electronic Instruments: Examines enclosure materials and internal insulation
used in measuring, monitoring, and control equipment.
-
Electrical Connectors and Terminal Blocks: Determines how insulating housings perform when
positioned close to locations where electrical arcing may occur.
-
Printed Circuit Board Assembly: Assesses circuit board substrates and component supports to
establish their susceptibility to flame propagation following a localized fault.
-
Insulating Materials and Engineering Plastics: Compares different material formulations
during development and incoming material quality inspections.
Standards
The QualiFlame™ NFT supports testing in accordance with established requirements for needle-flame apparatus
configuration, flame characteristics, and fire hazard evaluation.
-
IEC 60695-11-5: Fire Hazard Testing, Part 11-5: Test Flames, Needle-Flame Test Method,
Apparatus, Confirmatory Test Arrangement and Guidance.
-
UL 746A: Standard for Polymeric Materials, Short Term Property Evaluations.
Needle Flame Tester – QualiFlame™ NFT Key Features
The QualiFlame™ NFT is built to control the conditions that influence needle-flame test results, from burner
positioning and exposure duration to flame verification and temperature measurement.
-
Programmable PLC-Based Operation: A PLC controller coordinates the testing sequence through
a touchscreen interface, managing flame application and timing to minimize variations caused by manual
operation.
-
Three Burner Positioning Angles: The burner can be positioned at 0°, 20°, or 45°, allowing
laboratories to reproduce the flame orientation specified for different product evaluations.
-
Precision Stainless Steel Burner: A corrosion-resistant stainless steel burner with a 0.5 mm
bore produces the needle flame required by the test method while maintaining the specified burner geometry.
-
Flexible Flame Exposure Programming: Operators can set exposure periods anywhere from 0 s to
999.9 s, with 0.1 s resolution. This accommodates the commonly specified 30-second test as well as longer
exposure requirements.
-
Integrated Afterflame Monitoring: The instrument automatically measures and records how long
a specimen continues burning after flame removal. A manual stop function gives operators additional control
when the end of a burning event requires judgment.
-
12 mm Flame Height Verification: A dedicated measurement gauge allows operators to check
flame height before testing, helping ensure that every specimen receives the prescribed flame exposure.
-
Thermal Output Verification with Copper Block: The supplied Cu-ETP copper block provides a
measurable reference for checking burner performance. Its specified temperature rise must occur within 23.5
seconds to confirm that the flame meets the required thermal conditions.
-
Armored Type K Thermocouple: The 0.5 mm thermocouple withstands temperatures up to 1100°C
and provides rapid temperature response during timed burner calibration.
-
Standard Bottled Gas Operation: The burner accepts commercially available propane or butane
with a minimum purity of 95%, allowing laboratories to operate without installing a dedicated specialty gas
supply.
-
Integrated Draught-Free Test Chamber: With an internal volume exceeding 0.5 m³, the chamber
provides the space required for needle-flame testing under draught-free conditions without an additional
enclosure.
-
Temperature Measurement from 0°C to 1000°C: The instrument covers a broad measurement range
for burner calibration and specimen temperature monitoring, eliminating the need for a separate temperature
measurement instrument for these tasks.
-
English Interface with Language Options: The standard English-language display is suitable
for North American laboratory operations, while alternative language configurations can be provided upon
request.
-
Adaptable Electrical Configuration: The power supply can be customized to accommodate local
mains requirements when the standard 220 V configuration is unsuitable.
Theory and Method
The needle-flame test examines how materials respond when a localized fire originates from an electrical fault within
equipment. Such an event may begin at an arcing electrical contact or when insulation catches fire. The assessment
focuses on whether the initial flame remains confined to the affected area or causes nearby components to ignite and
sustain further burning. Unlike glow wire testing, which simulates excessive heating before ignition occurs, the
needle-flame method introduces an actual flame with controlled dimensions and thermal characteristics to evaluate
the specimen's response.
The test flame is generated by a needle burner featuring a 0.5 mm bore and supplied with propane or butane. Once
correctly adjusted, the burner produces a diffusion flame with a height of 12 mm. Maintaining this specified height
is essential because variations in flame dimensions affect the amount of heat transferred to the specimen. The
supplied flame-height gauge is therefore used to check the burner before each test series. The burner can also be
positioned at different angles to reproduce the required flame contact conditions. For example, a 45° orientation
creates a different pattern of heat exposure across the specimen surface compared with a vertically applied flame.
Burner heat output is verified using a drilled Cu-ETP copper block weighing 0.58 g, together with a 0.5 mm Type K
thermocouple. Under the specified flame conditions, the copper block must achieve the required temperature rise
within 23.5 seconds. Meeting this time requirement confirms that the burner delivers the prescribed thermal energy
and gives the laboratory measurable calibration evidence rather than relying solely on visual flame assessment.
Once flame verification is complete, the specimen is exposed for the duration required by the applicable testing
standard, typically 30 seconds. The evaluation then considers how long the specimen continues burning after the
flame is removed, whether fire spreads across the material, and whether burning particles or other material ignite
the specified material positioned beneath the specimen.
Needle Flame Tester - QualiFlame™ NFT Technical Specifications
| Parameter |
Specification |
| Model |
QualiFlame™ NFT |
| Needle Burner |
Ø0.5 mm ±0.1 mm inner bore, ≥9 mm outer diameter, ≥35 mm length, stainless steel |
| Burner Angle |
0°, 20°, 45° (flame height adjustable) |
| Thermocouple |
Ø0.5 mm Type K (Ni/Cr-Ni/Al), armored sheath rated to 1100°C |
| Flame Application Time |
0 s to 999.9 s ±0.1 s, adjustable (30 s typical) |
| Afterflame Timing |
0 s to 999.9 s ±0.1 s, automatic recording with manual timeout |
| Flame Height |
12 mm ±1 mm (measurement gauge supplied) |
| Flame Gas |
Propane or butane, 95% minimum purity (bottled lighter gas acceptable) |
| Temperature Measurement Range |
0°C to 1000°C |
| Flame Temperature Rise Requirement |
100°C ±2°C to 700°C ±3°C in 23.5 s ±1 s |
| Calibration Copper Block |
Ø4 mm ±0.01 mm, 0.58 g ±0.01 g before drilling, Cu-ETP |
| Test Process |
Automatic control, independent test cycle |
| Chamber Volume |
≥0.5 m³ |
| Controller |
PLC with integrated touch interface |
| Dimensions (L × W × H) |
1100 × 740 × 1250 mm |
| Voltage |
220 V, 48 Hz to 60 Hz, 800 VA (110V is also available) |
| Display Panel Language |
English (other languages available on request) |