Comparative Tracking Index Tester – QualiFlame™ CTI
Comparative Tracking Index (CTI) and Proof Tracking Index (PTI) Test Apparatus
The QualiFlame™ CTI evaluates the ability of solid electrical insulation materials to withstand surface tracking when
exposed to contaminated moisture under an applied electrical voltage. It establishes two important material
performance values: the Comparative Tracking Index (CTI) and the Proof Tracking Index (PTI). The instrument is
designed for testing environments across household appliance manufacturing, lighting equipment production,
electrical connector manufacturing, and material research laboratories.
During testing, two platinum electrodes make contact with the specimen while controlled electrolyte drops are
delivered into the space between them. The entire procedure takes place inside a fully enclosed test cabinet,
isolating the operator from energized components. This arrangement supports controlled testing while keeping direct
access to high-voltage parts restricted.
The specimen is positioned on a glass plate within an illuminated test chamber. Once in place, the electrodes descend
onto its upper surface. A supplied gauge block establishes the required 4.0 mm separation between the electrodes,
while the adjustable specimen table provides precise control over electrode positioning and the height from which
electrolyte drops fall. Before starting, the operator configures the test voltage, current limit, dripping
frequency, and total drop count through the control panel. Closing the chamber door allows the PLC-controlled
sequence to proceed without requiring access to the testing area.
The instrument accommodates test voltages from 100 V to 1000 V, with voltage displayed as an r.m.s. value and current
measurement held within a 1.5% tolerance. Its PLC-based touch control system manages the test sequence, records drop
counts, and identifies operating faults. If tracking failure continues for at least two seconds, an audible alarm
identifies the test endpoint. This automatic notification eliminates the need for continuous observation of the
current meter. Flaming ignition remains a separate observation, as the test method distinguishes it from electrical
tracking failure.
Comparative Tracking Index Tester – QualiFlame™ CTI Applications
The QualiFlame™ CTI is used to investigate how electrical insulation materials perform when exposed to
moisture-related surface contamination and electrical stress. Its CTI and PTI measurements help manufacturers
determine whether insulating components can maintain adequate electrical separation in their intended operating
environments. The tester is particularly relevant to product development, material selection, quality assurance, and
certification activities across the following industries:
-
Household Appliance Manufacturing: Supports insulation material selection during the design and
qualification of household electrical products. Manufacturers can use CTI results to establish appropriate
creepage distances for terminal assemblies, internal control components, and appliance enclosures, helping
assess the risk of surface discharge between energized parts.
-
Luminaire and Lighting Component Production: Helps lighting manufacturers evaluate insulation
materials used in lamp holders, LED driver housings, and structural electrical supports. Testing is particularly
relevant when these components must maintain electrical separation in environments where airborne dust and
moisture can contribute to surface contamination.
-
IT and Audio-Video Equipment: Assesses the ability of insulating components in power supplies,
adapters, and electronic equipment enclosures to resist conductive surface deterioration. The resulting
measurements provide material performance data for evaluating insulation suitability in compact electronic
assemblies.
-
Connectors and Terminal Blocks: Provides a basis for comparing the tracking resistance of
molded insulating housings used to separate neighboring electrical terminals. This is especially important in
connector designs where limited spacing between energized conductors makes the insulating surface a critical
part of electrical protection.
-
Switchgear and Circuit Protection: Enables manufacturers to examine the resistance of molded
insulation and internal barrier materials used in circuit breakers, contactors, and distribution boards. CTI
results help determine whether these materials can withstand contamination-related electrical stress while
maintaining separation between conductive components.
-
Thermoset and Thermoplastic Compounding: Assists material developers in evaluating how
different resin formulations and filler combinations influence electrical tracking resistance. Comparative
testing makes it possible to rank material grades according to their performance and identify suitable
formulations during product development.
-
Printed Circuit Board Laminates: Evaluates the tendency of insulating substrates to develop
conductive tracks under electrical stress, particularly in circuit board designs with closely spaced conductive
traces. The measurements help characterize the surface insulation properties of laminate materials.
-
Automotive and EV High-Voltage Components: Supports the qualification of insulating materials
used in battery packs, junction boxes, and charging assemblies. These components may encounter condensation
during operation, making resistance to moisture-assisted surface tracking an important material performance
consideration.
-
Certification and Third-Party Laboratories: Provides standardized CTI and PTI measurements for
independent material evaluation and product qualification. Laboratories can use the test results to document
insulation performance in technical datasheets and support product approval submissions.
Standards
The QualiFlame™ CTI supports the determination of comparative and proof tracking indices in accordance with the
applicable test method and electrical product safety standards. These standards reference tracking resistance as
part of evaluating the suitability of solid insulating materials in different applications.
-
IEC 60112 – Method for the Determination of the Proof and the Comparative Tracking Indices of Solid
Insulating Materials
-
IEC 60335-1 – Household and Similar Electrical Appliances, Safety, Part 1: General Requirements
-
IEC 60598-1 – Luminaires, Part 1: General Requirements and Tests
-
IEC 60950-1 – Information Technology Equipment, Safety, Part 1: General Requirements
Comparative Tracking Index Tester – QualiFlame™ CTI Key Features
The QualiFlame™ CTI combines automated electrical testing with precision positioning and specimen handling
provisions. Each component contributes to consistent test conditions, controlled electrolyte delivery, and reliable
identification of tracking failure.
-
Fully Enclosed Test Cabinet: Houses the energized testing assembly behind a closed door,
preventing direct operator exposure to the high-voltage electrodes while a test is underway.
-
PLC Control with Touch Interface: Coordinates voltage application, electrolyte drop
intervals, and drop counting through an automated testing sequence, reducing the need for manual operation.
-
99% Minimum Purity Platinum Electrodes: Provide resistance to oxidation and material
erosion, reducing electrode-related variations that could affect results over repeated testing cycles.
-
Chisel-Edge Electrode Geometry: Features the specified 30° ±2° edge angle to establish
consistent electrode contact geometry across different specimens.
-
Supplied Gauge Block: Provides a practical reference for setting the required 4.0 mm
distance between electrodes, allowing accurate and repeatable setup.
-
Adjustable Table Height: Allows precise positioning of the specimen to establish the
required electrode angle and electrolyte drop height for each test arrangement.
-
Programmable Voltage and Limiting Current: Offers an adjustable voltage range of 100 V to
1000 V, supporting both PTI proof testing and complete CTI evaluations on one instrument.
-
Two-Second Tracking Failure Alarm: Automatically alerts the operator when the tracking
failure condition persists for two seconds or longer, eliminating the need for uninterrupted current meter
monitoring.
-
Dual Dripping Nozzle Sizes: Accommodates the different electrolyte drop volumes specified
for Solution A and Solution B testing.
-
Illuminated Chamber: Provides visibility throughout the test, allowing observation of
scintillation, surface deterioration, and any flaming ignition.
-
Calibrated 1.00 N Electrode Force: Applies the specified electrode contact force to
establish consistent testing pressure and support comparable results between laboratories.
-
Customizable Power Supply and Interface Language: Offers configuration options for local
electrical supply requirements and operator language preferences.
Theory and Method
Electrical surface tracking develops through a sequence of localized deterioration rather than an immediate breakdown
of the insulation. The process begins when contaminated water forms a partially conductive path across the surface
between two energized points. As leakage current passes through this moisture film, localized heating causes water
to evaporate unevenly. The resulting dry regions, known as dry bands, interrupt the conductive film and concentrate
the applied voltage across the remaining gaps.
Electrical discharges then develop across these dry bands, producing small surface arcs referred to as
scintillations. Repeated discharges subject the polymer surface to pyrolysis, leaving behind carbonized deposits.
Unlike the original insulating material, this carbon residue provides a more conductive path across the specimen.
As carbon deposits accumulate, the effective insulating distance between the electrodes becomes progressively
shorter. This allows further leakage current and arcing to develop, accelerating the deterioration process until the
material reaches its failure condition.
The CTI test reproduces this degradation mechanism under controlled laboratory conditions. Two platinum electrodes
with chisel-shaped edges are positioned 4.0 mm apart and apply a contact force of 1.00 N to the specimen.
Electrolyte drops are then delivered into the gap at predetermined intervals to create the conditions required for
tracking evaluation.
Two electrolyte formulations are available for the test. Solution A consists of a dilute ammonium chloride
electrolyte. Solution B contains an additional wetting agent that reduces surface tension, allowing the liquid to
spread more readily over hydrophobic insulating surfaces. This increased surface coverage creates more demanding
test conditions.
A specimen reaches the failure endpoint when its leakage current exceeds the specified method threshold continuously
for at least two seconds or when ignition occurs. These conditions establish the basis for evaluating its tracking
resistance.
The Proof Tracking Index (PTI) identifies whether an insulating material can withstand 50 consecutive electrolyte
drops at a selected test voltage without reaching the failure condition. The Comparative Tracking Index (CTI), in
contrast, establishes the highest voltage at which the material successfully withstands the 50-drop test. This value
provides an important reference for determining the creepage distances permitted in electrical product design.
Electrode material selection is also critical to test accuracy. Platinum is used because it resists oxidation and
minimizes the possibility of electrode-derived oxides contaminating the specimen surface. A more reactive electrode
material could introduce additional oxides into the developing track, potentially influencing the observed tracking
behavior and compromising the consistency of test results.
Comparative Tracking Index Tester - QualiFlame™ CTI Technical Specifications
| Parameter |
Specification |
| Model |
QualiFlame™ CTI |
| Controller |
PLC with integrated touch interface |
| Control Language |
English (other languages available on request) |
| Electrode Material |
Platinum, 99% minimum purity |
| Electrode Separation |
4.0 mm ±0.1 mm |
| Electrode Angle |
60° ±5° |
| Electrode Width |
5 mm ±0.1 mm |
| Electrode Thickness |
2 mm ±0.1 mm |
| Electrode Length |
>12 mm |
| Electrode End Chisel-Edge Angle |
30° ±2°, platinum |
| Electrode Pressure |
1.00 N ±0.05 N |
| Test Voltage |
100 V to 1000 V, adjustable, r.m.s. display |
| Test Current |
r.m.s. display, tolerance 1.5% |
| Dripping Nozzle, Solution A |
0.9 mm to 1.2 mm outer diameter |
| Dripping Nozzle, Solution B |
0.9 mm to 3.45 mm outer diameter |
| Dimensions (L × W × H) |
1100 × 740 × 1250 mm |
| Voltage |
220 V, 50 Hz to 60 Hz, single phase, 800 W (110V is also available) |