{"product_id":"ball-pressure-test-apparatus","title":"Ball Pressure Test Apparatus","description":"\u003ch2\u003eBall Pressure Test Apparatus – QualiHSA™ BPT\u003c\/h2\u003e\n\u003cp\u003eThe QualiHSA™ BPT Ball Pressure Test Apparatus evaluates the heat resistance of non-metallic materials by pressing a\n    5 mm steel ball against a heated test specimen under a 20 N load. The resulting indentation indicates how well the\n    material retains its shape when exposed to elevated temperatures. This method is commonly used by compliance\n    laboratories to assess electrical equipment enclosures, insulating supports, and other non-metallic components.\n    Testing takes place inside a heating cabinet maintained at the temperature specified by the applicable product\n    standard. After the specimen has cooled, the operator measures the indentation. A diameter of no more than 2.0 mm\n    indicates compliance with the acceptance criterion.\u003c\/p\u003e\n\u003cp\u003eBoth the loading assembly and specimen support are manufactured from stainless steel to maintain dimensional\n    stability during repeated exposure to high test temperatures. The specimen rests on a solid stainless steel cylinder\n    measuring 50 mm in diameter and 100 mm in length. Its substantial thermal mass helps maintain consistent specimen\n    temperature throughout the test. The complete applied load, including the weight of the steel ball, is maintained at\n    20 N ± 0.2 N. Since the QualiHSA™ BPT operates with a separate heating cabinet, laboratories can use an existing\n    oven rather than requiring a dedicated integrated heating system.\u003c\/p\u003e\n\u003cp\u003eManufacturers of plugs, sockets, household appliances, luminaires, and IT equipment regularly encounter ball pressure\n    testing requirements as part of their product safety assessments. Multiple IEC product standards specify this\n    method, with the required test temperature determined by the particular product and component under evaluation. The\n    same apparatus can accommodate these different testing programs because the ball dimensions, loading force, and\n    specimen support configuration remain unchanged. Laboratories only need to adjust the heating cabinet temperature\n    according to the applicable standard.\u003c\/p\u003e\n\u003ch2\u003eBall Pressure Test Apparatus – QualiHSA™ BPT Applications\u003c\/h2\u003e\n\u003cp\u003eThe QualiHSA™ BPT is suitable for evaluating non-metallic components used in electrotechnical equipment, including\n    complete subassemblies and individual parts. It also supports testing of solid electrical insulating materials,\n    excluding ceramics. Typical test specimens include molded components responsible for supporting electrical\n    connections, protecting internal assemblies, or maintaining separation between energized conductors.\u003c\/p\u003e\n\u003cul\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003ePlugs, Socket-Outlets, and Appliance Couplers\u003c\/strong\u003e: Determines whether housings and insulating\n            barriers retain their structural integrity under heat exposure, as required by IEC 60884-1 and IEC 60320-1.\n        \u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eHousehold Appliance Manufacturing\u003c\/strong\u003e: Assesses appliance enclosures and terminal blocks to\n            establish whether elevated temperatures could soften their materials and cause energized components to shift\n            from their intended positions.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eLuminaire Production\u003c\/strong\u003e: Evaluates lampholders, driver housings, and insulating spacers at the\n            test temperatures specified in the relevant luminaire standards.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eIT and Consumer Electronics\u003c\/strong\u003e: Checks power supply enclosures, connector bodies, and internal\n            support components for permanent deformation when exposed to operating temperatures.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eTerminal Block and Connector Suppliers\u003c\/strong\u003e: Verifies that thermoplastic housings maintain their\n            shape when subjected to prolonged heat exposure and clamping forces.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eSolid Insulating Material Development\u003c\/strong\u003e: Provides a consistent method for comparing different\n            thermoplastic and thermoset materials by measuring indentation diameter at a specified temperature.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eSafety Certification Laboratories\u003c\/strong\u003e: Supports routine heat resistance assessments across\n            multiple IEC product categories using a single testing fixture.\u003c\/p\u003e\n    \u003c\/li\u003e\n    \u003cli\u003e\n        \u003cp\u003e\u003cstrong\u003eIncoming Quality Control\u003c\/strong\u003e: Enables manufacturers to examine batches of supplier-provided\n            molded components for heat resistance before releasing them for production assembly.\u003c\/p\u003e\n    \u003c\/li\u003e\n\u003c\/ul\u003e\n\n\n\n\u003ch2\u003eStandards\u003c\/h2\u003e\n\u003cp\u003eThe QualiHSA™ BPT supports ball pressure testing requirements specified across several IEC standards covering\n    electrical appliances, connection accessories, lighting equipment, and IT products. These standards establish the\n    requirements for assessing the resistance of non-metallic insulating components to deformation under elevated\n    temperatures. The apparatus provides the specified 5 mm ball, 20 N loading force, and specimen support arrangement,\n    while the applicable product standard determines the temperature at which each component must be evaluated.\u003c\/p\u003e\n\u003cp\u003eThe relevant standards include:\u003c\/p\u003e\n\u003cul\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60695-10-2 – Ball Pressure Test Method:\u003c\/strong\u003e Specifies the test method for evaluating the\n            resistance of materials to abnormal heat through controlled ball pressure loading.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60335-1 – Household and Similar Electrical Appliances:\u003c\/strong\u003e Covers general safety\n            requirements for electrical appliances, including relevant heat resistance assessments of non-metallic\n            components.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60598-1 – Luminaires:\u003c\/strong\u003e Establishes general requirements and testing provisions for\n            lighting equipment.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60884-1 – Plugs and Socket-Outlets:\u003c\/strong\u003e Defines general requirements for household and\n            similar electrical plugs and socket-outlets.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60320-1 – Appliance Couplers:\u003c\/strong\u003e Addresses general requirements for appliance couplers used\n            in household and similar applications.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eIEC 60950-1 – Information Technology Equipment:\u003c\/strong\u003e Covers general safety requirements for IT\n            equipment.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cp\u003eThe required test temperature must be selected according to the applicable product standard and the component being\n    assessed.\u003c\/p\u003e\n\n\n\n\n\n\n\u003ch2\u003eBall Pressure Test Apparatus – QualiHSA™ BPT Key Features\u003c\/h2\u003e\n\u003cp\u003eThe QualiHSA™ BPT is built around a mechanically controlled loading arrangement that delivers the specified test\n    conditions without complicated operating systems. Its stainless steel construction, defined specimen support\n    dimensions, and dead-weight loading mechanism provide the essential components for repeatable ball pressure testing.\n    Each design element serves a specific purpose, from maintaining the applied force to preserving specimen temperature\n    during exposure.\u003c\/p\u003e\n\u003cul\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003ePrecisely Defined Pressure Ball Geometry\u003c\/strong\u003e: Uses a 5 mm diameter steel ball with an R2.5 mm\n            spherical surface, providing the standardized contact geometry required for consistent indentation testing.\n        \u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eControlled Dead-Weight Loading\u003c\/strong\u003e: Applies a total force of 20 N ± 0.2 N, including the weight\n            of the pressure ball. The mechanical loading arrangement avoids spring-based force application.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eThermally Stable Specimen Support\u003c\/strong\u003e: Features a solid stainless steel cylinder measuring 50 mm\n            in diameter and 100 mm in length. Its substantial mass helps maintain the specimen at the required\n            temperature throughout the test.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eStainless Steel Loading and Support Components\u003c\/strong\u003e: Both assemblies are manufactured from\n            stainless steel to withstand repeated high-temperature exposure while resisting oxidation and dimensional\n            changes.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eCompatible with Separate Heating Cabinets\u003c\/strong\u003e: Operates independently of the heating system,\n            allowing laboratories to select the required test temperature using an existing heating cabinet with\n            suitable capacity.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eSimple Mechanical Test Arrangement\u003c\/strong\u003e: Allows operators to prepare the fixture, position the\n            heated specimen, and apply the pressure ball without electrical connections or electronic controls.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eVersatility Across IEC Testing Programs\u003c\/strong\u003e: Accommodates ball pressure assessments for\n            electrical appliances, plugs, socket-outlets, appliance couplers, luminaires, and IT equipment without\n            changing the fixture geometry.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n    \u003cli\u003e\n\n\n        \u003cp\u003e\u003cstrong\u003eMinimal Servicing Requirements\u003c\/strong\u003e: Its construction contains no motors, sensors, or electronic\n            control components. Routine maintenance primarily involves checking the pressure ball's condition and\n            verifying the applied loading force.\u003c\/p\u003e\n\n\n    \u003c\/li\u003e\n\u003c\/ul\u003e\n\n\n\n\u003ch2\u003eTheory and Method\u003c\/h2\u003e\n\u003cp\u003eWhen thermoplastic materials are exposed to increasing temperatures, their ability to resist deformation\n    progressively decreases as they approach their softening range. This change does not necessarily occur at a clearly\n    defined point. Instead, a material that remains rigid under normal room conditions may gradually deform when\n    subjected to a constant mechanical force at an elevated temperature. The ball pressure test provides a measurable\n    way to assess this behavior by maintaining a defined load and contact geometry for a specified period, then\n    examining the permanent indentation left in the specimen. This evaluation is particularly important for electrical\n    insulating components, where heat-induced deformation can alter the position of supporting structures, reduce\n    creepage distances, or bring energized conductors closer to surrounding components.\u003c\/p\u003e\n\u003cp\u003eDuring testing, the 5 mm steel ball applies a 20 N force over a relatively small contact area, initially\n    concentrating the load into a contact circle considerably smaller than one millimeter. This produces a high\n    localized stress compared with the pressure generated by a flat-ended loading tool. As the ball penetrates the\n    material, its curved surface progressively increases the contact area. The resulting reduction in localized stress\n    limits further penetration and helps prevent uncontrolled indentation. Maintaining the specified force for 60\n    minutes allows the material's viscoelastic response to develop, including time-dependent creep under sustained\n    loading. Once the exposure period is complete, cooling the specimen in water stabilizes the resulting deformation\n    before elastic recovery can significantly influence the measurement.\u003c\/p\u003e\n\u003cp\u003eReliable results also require the specimen to remain at the prescribed temperature throughout the loading period. The\n    solid stainless steel support cylinder provides the thermal mass needed to maintain the temperature around the\n    contact area while the pressure ball acts on the specimen. Its stainless steel construction also resists surface\n    scaling that could gradually affect flatness and introduce inconsistencies between tests. Before testing begins, the\n    support and loading assembly are heated inside the cabinet so that introducing the specimen does not create an\n    unwanted temperature fluctuation. Following the prescribed cooling process, the operator measures the indentation\n    without unnecessary delay. A magnifier and calibrated cross-travel stage are used to determine the indentation\n    diameter, providing the measurement needed to evaluate the specimen against the applicable acceptance limit.\u003c\/p\u003e\n\u003ch2\u003eBall Pressure Test Apparatus - QualiHSA™ BPT Technical Specifications\u003c\/h2\u003e\n\n\u003ctable class=\"table table-style-align-center\" align=\"center\" border=\"1\" cellspacing=\"0\" cellpadding=\"0\"\u003e\n    \u003cthead\u003e\n        \u003ctr\u003e\n            \u003cth\u003eParameter\u003c\/th\u003e\n            \u003cth\u003eSpecification\u003c\/th\u003e\n        \u003c\/tr\u003e\n    \u003c\/thead\u003e\n    \u003ctbody\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eBall Diameter\u003c\/td\u003e\n            \u003ctd\u003e5 mm (R2.5 mm)\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eTotal Test Force\u003c\/td\u003e\n            \u003ctd\u003e20 N ± 0.2 N, including the mass of the ball\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eSpecimen Support\u003c\/td\u003e\n            \u003ctd\u003eSolid stainless steel cylinder, 50 mm diameter × 100 mm long\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eMaterials of Construction\u003c\/td\u003e\n            \u003ctd\u003eStainless steel throughout\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eHeating Source\u003c\/td\u003e\n            \u003ctd\u003eSeparate heating cabinet or oven, not supplied\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eTest Temperature\u003c\/td\u003e\n            \u003ctd\u003eSet by the applicable product standard\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eAcceptance Criterion\u003c\/td\u003e\n            \u003ctd\u003eIndentation diameter not exceeding 2.0 mm\u003c\/td\u003e\n        \u003c\/tr\u003e\n    \u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003ch3\u003eTest Sequence\u003c\/h3\u003e\n\u003ctable class=\"table table-style-align-center\" align=\"center\" border=\"1\" cellspacing=\"0\" cellpadding=\"0\"\u003e\n    \u003cthead\u003e\n        \u003ctr\u003e\n            \u003cth\u003eStep\u003c\/th\u003e\n            \u003cth\u003eParameter\u003c\/th\u003e\n        \u003c\/tr\u003e\n    \u003c\/thead\u003e\n    \u003ctbody\u003e\n        \u003ctr\u003e\n            \u003ctd\u003ePreheat\u003c\/td\u003e\n            \u003ctd\u003eSupport and loading device brought to test temperature in the heating cabinet before the specimen is\n                introduced\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eHold at temperature\u003c\/td\u003e\n            \u003ctd\u003e60 minutes, +2 \/ 0 minutes\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eTransfer to water\u003c\/td\u003e\n            \u003ctd\u003eWithin 10 seconds of removing the load\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eCooling\u003c\/td\u003e\n            \u003ctd\u003eSpecimen cooled in water at 20 °C ± 5 °C until it returns to approximately room temperature\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eMeasurement\u003c\/td\u003e\n            \u003ctd\u003eIndentation diameter read within 3 minutes of removal from the water\u003c\/td\u003e\n        \u003c\/tr\u003e\n        \u003ctr\u003e\n            \u003ctd\u003eResult\u003c\/td\u003e\n            \u003ctd\u003ePass when the indentation diameter does not exceed 2.0 mm\u003c\/td\u003e\n        \u003c\/tr\u003e\n    \u003c\/tbody\u003e\n\u003c\/table\u003e","brand":"Qualitest USA LC","offers":[{"title":"Default Title","offer_id":53636895703351,"sku":null,"price":100.0,"currency_code":"USD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0786\/1859\/1543\/files\/ball-pressure-test-apparatus4814.webp?v=1794892895","url":"https:\/\/qualitest.us\/fr\/products\/ball-pressure-test-apparatus","provider":"Qualitest USA LC","version":"1.0","type":"link"}